262 T-Slot Transducer Removal and Installation Procedures
Section 7: Maintenance and troubleshooting Reference, Installation, and Operations Manual
August 2014 3-9000-743 Rev U
7.2 T-Slot Transducer Removal and Installation
Procedures
The T-Slot transducer assembly is an improvement to the direct mount transducer. It offers
improved transducer alignment and superior acoustic isolation between the transducer and the
meter housing. The net result is improved performance and stability. The assembly is used on
both SeniorSonic™ and JuniorSonic™ Meters and is line pressure vented.
7.2.1 Removal with Extractor Tool
To replace transducers under pressure, contact Daniel Customer Service. Only trained and
certified Daniel Service Technicians are allowed to perform maintenance operations on Daniel
SeniorSonic Meters using the extractor tool. The contact information is located in the Preface
and on the back page of this manual. You can access the request form from the Daniel
Measurement and Control web site; Daniel Life Cycle Services page:
http://www2.emersonprocess.com/en-US/brands/daniel/support-services/Measurement-
Analytical/Pages/MaintenanceRepairServices.aspx
Refer to the following images to determine the transducer assembly installed in your meter and
provide this information to the Customer Service Representative.
Figure 7-1 J-Mount Transducer Assembly
A. Transducer cable (max. length 15 ft.)
B. Transducer cable nut and chordset
C. Mount
D.
Transducer assembly, stalk and holder,
A.
B.
C.
D.